Strange Observations – Part 38

Continuing in a series of strange things that I have seen while consulting at hospitals…

This hospital wanted to ‘fence off’ access to the MRI when it was not staffed. Apparently, they did not have a side-hinged swinging door assembly available, so they went with this removable gate.

The problem is, the gate is not side-hinged, and swinging and on top of that they put a padlock on the gate so anyone left in the room would be locked in.

Their reply was, they would ensure that no one is in the MRI area when they set up the gate. Sorry, but the LSC does not allow for that. The gate had to go.